Type: TCZ 4022 S -2K
Description: Climate simulation chamber, air-cooled
System Design:
The system consists of the following main assemblies:
- Walk-in, insulated chamber
- Air handling room
- Humidification unit
Ibsdjvx S Hhspfx Agusc
- Machine room
- Control cabinet
Technical Data:
Performance Range for Temperature Testing:
- Temperature range: -40 to +140°C
- Temperature stability, in supply air (over time): ≤ ±0.5°C
(without thermal test chamber load, according to DIN 50011 T13 / EN 60068)
- Temperature stability, in supply air (spatial): ≤ ±1.0°C
(without thermal test chamber load, according to DIN 50011 / EN 60068)
- Heating rate (according to DIN 50011): approx. 2.0 K/min
(without thermal test chamber load, according to DIN 50011 / EN 60068), without test object, within -30°C to +85°C
- Cooling rate (according to DIN 50011): approx. 2.0 K/min
(without thermal test chamber load, according to DIN 50011 / EN 60068), without test object, within -30°C to +85°C
Performance Range for Climate Testing:
- Temperature range: +10 to +90°C
- Humidity operating range: 10 to approx. 95% RH
(without thermal test chamber load)
- Dew point temperature range: +5 to +89°C
(without thermal test chamber load)
- Humidity stability (in supply air, over time): ≤ ±2.5% RH
(measured in supply air)
Performance Range for Climate Testing with Heat Compensation:
- Temperature range: +15 to +80°C
- Humidity: 15 to approx. 80% RH
(without thermal test chamber load, with max. 16 kW heat compensation)
- Dew point temperature range: +10 to +74°C
(without thermal test chamber load)
Dimensions:
- Test chamber (W x D x H): approx. 2260 x 4648 x 2260 mm
- External dimensions (W x D x H): approx. 2660 x 6460 x 2800 mm
- Clear door opening test chamber door (W x H): 2260 x 2260 mm
- Climate module KB 127 (W x D x H): 1700 x 800 x 1400 mm
- Control cabinet (W x D x H): 800 x 500 x 2000 mm
Design:
- Air guidance: Supply air via perforated sheet at rear wall, ceiling area
Exhaust air via perforated sheet at rear wall, floor area
- Temperature control: Direct cooling via evaporation of refrigerant in a finned heat exchanger; Heating with electric resistance heater
- Dehumidification: Direct dehumidification by evaporating refrigerant in a heat exchanger
- Internal lining: Stainless steel
- External cladding: galvanized, powder-coated steel sheet
- Floor construction: Stainless steel, floor load capacity 15,000 N/m²
Wheel load (contact area ≥4 cm²): 1,500 N/wheel
- Door: Double-leaf, stainless steel, external painted, with viewing window (W x H) 1000 x 1000 mm
- Test chamber illumination: 4 spotlights at 500 W each
- Insulation: Double-layer, 80 mm PUR foam, 120 mm mineral wool
Refrigeration Unit (1 piece): Bitzer 6G-30.2Y
- Refrigerant: R449A
- Refrigeration oil: SEZ 32
- Steam humidifier (1 piece): EDB 40
Color scheme:
- Chamber: RAL9010, white
- Door: RAL9010, white
- Machine unit: RAL7032, pebble grey
- Control cabinet: RAL7032, pebble grey
Weights:
- Total system: approx. 4,900 kg
- Machine set KB 127: approx. 750 kg
- Control cabinet: approx. 120 kg
- Humidification unit: approx. 64 kg
Connections:
- Nominal voltage: 400 V +6% / -10%, 3/PE/N, 50 Hz
- Nominal power: 95 kW
- Nominal current: 150 A
Inzercia bola automaticky preložená a vyskytli sa nejaké chyby prekladu.
Type: TCZ 4022 S -2K
Description: Climate simulation chamber, air-cooled
System Design:
The system consists of the following main assemblies:
- Walk-in, insulated chamber
- Air handling room
- Humidification unit
Ibsdjvx S Hhspfx Agusc
- Machine room
- Control cabinet
Technical Data:
Performance Range for Temperature Testing:
- Temperature range: -40 to +140°C
- Temperature stability, in supply air (over time): ≤ ±0.5°C
(without thermal test chamber load, according to DIN 50011 T13 / EN 60068)
- Temperature stability, in supply air (spatial): ≤ ±1.0°C
(without thermal test chamber load, according to DIN 50011 / EN 60068)
- Heating rate (according to DIN 50011): approx. 2.0 K/min
(without thermal test chamber load, according to DIN 50011 / EN 60068), without test object, within -30°C to +85°C
- Cooling rate (according to DIN 50011): approx. 2.0 K/min
(without thermal test chamber load, according to DIN 50011 / EN 60068), without test object, within -30°C to +85°C
Performance Range for Climate Testing:
- Temperature range: +10 to +90°C
- Humidity operating range: 10 to approx. 95% RH
(without thermal test chamber load)
- Dew point temperature range: +5 to +89°C
(without thermal test chamber load)
- Humidity stability (in supply air, over time): ≤ ±2.5% RH
(measured in supply air)
Performance Range for Climate Testing with Heat Compensation:
- Temperature range: +15 to +80°C
- Humidity: 15 to approx. 80% RH
(without thermal test chamber load, with max. 16 kW heat compensation)
- Dew point temperature range: +10 to +74°C
(without thermal test chamber load)
Dimensions:
- Test chamber (W x D x H): approx. 2260 x 4648 x 2260 mm
- External dimensions (W x D x H): approx. 2660 x 6460 x 2800 mm
- Clear door opening test chamber door (W x H): 2260 x 2260 mm
- Climate module KB 127 (W x D x H): 1700 x 800 x 1400 mm
- Control cabinet (W x D x H): 800 x 500 x 2000 mm
Design:
- Air guidance: Supply air via perforated sheet at rear wall, ceiling area
Exhaust air via perforated sheet at rear wall, floor area
- Temperature control: Direct cooling via evaporation of refrigerant in a finned heat exchanger; Heating with electric resistance heater
- Dehumidification: Direct dehumidification by evaporating refrigerant in a heat exchanger
- Internal lining: Stainless steel
- External cladding: galvanized, powder-coated steel sheet
- Floor construction: Stainless steel, floor load capacity 15,000 N/m²
Wheel load (contact area ≥4 cm²): 1,500 N/wheel
- Door: Double-leaf, stainless steel, external painted, with viewing window (W x H) 1000 x 1000 mm
- Test chamber illumination: 4 spotlights at 500 W each
- Insulation: Double-layer, 80 mm PUR foam, 120 mm mineral wool
Refrigeration Unit (1 piece): Bitzer 6G-30.2Y
- Refrigerant: R449A
- Refrigeration oil: SEZ 32
- Steam humidifier (1 piece): EDB 40
Color scheme:
- Chamber: RAL9010, white
- Door: RAL9010, white
- Machine unit: RAL7032, pebble grey
- Control cabinet: RAL7032, pebble grey
Weights:
- Total system: approx. 4,900 kg
- Machine set KB 127: approx. 750 kg
- Control cabinet: approx. 120 kg
- Humidification unit: approx. 64 kg
Connections:
- Nominal voltage: 400 V +6% / -10%, 3/PE/N, 50 Hz
- Nominal power: 95 kW
- Nominal current: 150 A